DWG-5088 Industrial online Sodium Meter
DWG-5088 Industrial Sodium Meter is a brand new continuous monitoring instrument for micro-sodium ions at
ppb level. With professional ppb level measuring electrode, automatic constant-voltage constant-current fluid line
system and stable and efficient basification system, it provides stable and accurate measurement. It can be used for
continuous monitoring over sodium ions in water and solution in thermal power stations, chemical industry,
chemical fertilizer, metallurgy, environmental protection, pharmacy, biochemical engineering, foodstuff, running
water supply and many other industries.
Features
LCD display in English, menu in English and notepad in English.
Computer-based: Multi-parameter measurement, temperature compensation and automatic range conversion are
achieved thanks to the high-performance CPU chip, high-precision AD conversion technology and SMT pasting
technology, resulting in high precision and repeatability.
High reliability: Single-board structure, touch keys, no switch knob or potentiometer. Rapid response, accurate
measurement and high stability.
Automatic constant-voltage constant-current liquid line system: Automatic compensation for flow and pressure of
water sample.
Alarm: Isolated alarm signal output, discretionary setting of upper and lower thresholds for alarming, and lagged
cancellation of alarming.
Network function: Isolated current output and RS485 Communication Interface.
History curve: It can continuously record data for a month, with a point for each five minutes.
Notepad function: Recording 200 messages.
Technical Indexes
1.Measuring range : 0 ~ 100ug / L, 0 ~ 23000mg /L
Resolution : 0.1 μg / L, 0.01mg/L0.00pNa-8.00pNa
Resolution : 0.01pNa0 ~ 60 ºC Resolution : 0.1 ºC
2.The basic error : ± 2.5%, ± 0.3 ºC temperature
3.Automatic temperature compensation range : 0 ~ 60 ºC, 25 ºC basis
4.The electronic unit temperature compensation error : ± 2.5%
5.The electronic unit repeatability error : ± 2.5% of reading
6. Stability: reading ± 2.5% / 24h
7.The input current : ≤ 2 x 10-12A Tested water samples : 0 ~ 60 ºC, 0.3MPa
8.The clock accuracy: ± 1 minute / month
9.The output current error : ≤ ± 1% FS
10.Data storage Quantity: 1 month ( 1:00 / 5 minutes )
11.Alarm normally open contacts :AC 250V, 7A
12.Power supply :AC220V ± 10%, 50 ± 1Hz
13.Isolated output : 0 ~ 10mA( load <1.5kΩ), 4 ~ 20mA ( load <750Ω)
14.Dimensions: 440 ( W) * 770 (H ) * 234 (D ) mm, hole size: 390 ( W ) * 650 ( H) mm
Positioning holes : 280 ( W ) * 730 (H ) mm, pore size : ¢ 12, four-hole distribution
Alarming relay: AC220V, 3A, isolated output of alarming signals
Aperture : ¢ 12, four-hole distribution ( unless otherwise noted , the products in accordance with the size of the
opening hole )
15. Weight : 20kg
16.Working conditions: ambient temperature: 0-60ºC; relative humidity <85%