Still deciding? Get samples of $ !
Request Sample
|
| Customization: | Available |
|---|---|
| After-sales Service: | Support |
| Warranty: | 1year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
| 1. Measuring range |
| Residual chlorine: 0-20.00mg/L, Resolution: 0.01mg/L; |
| HOCL: 0-10.00mg/L Resolution: 0.01mg/L; |
| pH value: 0 - 14.00pH Resolution: 0.01pH; |
| Temperature: 0- 99.9 ºC Resolution: 0.1 ºC |
| 2. Accuracy: Residual chlorine: ± 2% or ± 0.035mg / L, take the bigger; |
| HOCL: ± 2% or ± 0.035mg / L, take the bigger; |
| pH value: ± 0.05pH |
| Temperature: ± 0.5 ºC (0 ~ 60.0 ºC); |
| 3. Sample temperature: 0 ~ 60.0 ºC, 0.6MPa; |
| 4. Sample flow rate: 200 ~250 mL/1min Automatic and Adjustable |
| 5. Minimum detection limit: 0.01mg / L |
| 6. Isolated current output: 4~20 mA(load <750Ω) |
| 7. High and low alarm relays: AC220V, 7A; |
| hysteresis 0- 5.00mg / L, arbitrary regulation |
| 8. RS485 communication interface (optional) |
| It can be convenient to computer monitoring and communication |
| 9. The data storage capacity: 1 month (1 point/5 minutes) |
| 10. Power Supply: AC220V ± 22V, 50Hz ± 1Hz; DC24V (optional). |
| 11. Protection grade: IP65 |
| 12. Overall dimension: 146 (length) x 146 (width) x 108 (depth) mm; |
| dimension of the hole: 138 x 138mm |
| Note: The wall installation may be ok, please specify when ordering. |
| 13. Weight: Secondary Instrument: 0.8kg, flow cell with residual chlorine, |
| pH electrode weight: 2.5kg; |
| 14. Working Conditions: ambient temperature: 0 ~ 60 ºC; relative humidity <85%; |
| 15.Adopt the flow-through installation, inlet and outlet diameter at Φ10. |
| Note: The user can select two-way current output; one way of residual chlorine corresponds to current output, and the other way of optional pH or temperature corresponds to the current output. The ex-factory standard configuration is that one way of residual chlorine corresponds to the current output. Please specify when ordering. |