DOG-209FA Industrial Dissolved Oxygen Sensor
Features
DOG-209FA type oxygen electrode improved from the previously dissolved oxygen electrode, change diaphragm into a grit
mesh metal membrane, with high stability and stress resistant, can be used in a more harsh environment, maintenance
volume is smaller, suitable for urban sewage treatment, industrial wastewater treatment, aquaculture and environmental
monitoring and other fields of continuous measurement of dissolved oxygen.
Technical Indexes
Ultra resistant to pressure (0.6Mpa) wallop, imported (grit mesh metal membrane) |
Up thread: M32 * 2.0 |
Measuring range: 0-20mg / L |
Measuring principle: Current type sensor (polarographic electrode) |
Breathable membrane thickness: 100μm |
Electrode shell material: PVC or 316L stainless steel |
Temperature compensation resistance: Pt100, Pt1000, 22K, 2.252K, etc. |
Sensor life:> 2 years |
Cable length: 5m |
Detection limit: 0.01 mg / L (20 ºC) |
Measurement limit: 40 mg / L |
Response time: 2min (90%, 20 ºC) |
Polarization time: 60min |
Minimum flow rate: 2.5cm / s |
Drift: <2% / month |
Measurement error: <± 0.01 mg / L |
Output Current: 50-80nA/0.1 mg / L Note: Maximum current 3.5uA |
Polarization voltage: 0.7V |
Zero oxygen: <0.01 mg / L |
Calibration interval:> 60 days |
Measured water temperature :0-60 ºC |
DOG-209F Industrial Dissolved Oxygen Sensor
Features
DOG-209F Dissolved Oxygen electrode has high stability and reliability, which can be used in the harsh environment;
it demands for less maintenance; it is suitable for continuous measurement of the dissolved oxygen in the fields of urban
sewage treatment, industrial waste water treatment, aquaculture, environmental monitoring etc.
Technical Indexes
Measuring range: 0-20mg/L |
Measuring principle: Current sensor (Polarographic electrode) |
Permeable membrane thickness: 50 um |
Electrode shell material: U PVC or 31 6L stainless steel |
Temperature compensation resistor: Ptl00, Ptl000, 22K, 2.252K etc. |
Sensor life: >2 years |
Cable length: 5m |
Detection lower limit: 0.01 mg/L (20ºC) |
Measurement upper limit: 40mg/L |
Response time: 3min (90%, 20ºC) |
Polarization time: 60min |
Minimum flow rate: 2.5cm/s |
Drift: <2%/month |
Measurement error: <± 0.1mg/I |
Output current: 50~80nA/0.1mg/L Note: Maximum Current 3.5uA |
Polarization voltage: 0.7V |
Zero oxygen: <0.1 mg / L (5min) |
Calibration intervals: >60 days |
Measured water temperature: 0-60ºC |