DOG-209FA Industrial Dissolved Oxygen Sensor
Features
DOG-209FA type oxygen electrode improved from the previously dissolved oxygen electrode, change diaphragm into a grit
mesh metal membrane, with high stability and stress resistant, can be used in a more harsh environment, maintenance
volume is smaller, suitable for urban sewage treatment, industrial wastewater treatment, aquaculture and environmental
monitoring and other fields of continuous measurement of dissolved oxygen.
Technical Indexes
| Ultra resistant to pressure (0.6Mpa) wallop, imported (grit mesh metal membrane) |
| Up thread: M32 * 2.0 |
Measuring range: 0-20mg / L |
| Measuring principle: Current type sensor (polarographic electrode) |
| Breathable membrane thickness: 100μm |
| Electrode shell material: PVC or 316L stainless steel |
| Temperature compensation resistance: Pt100, Pt1000, 22K, 2.252K, etc. |
| Sensor life:> 2 years |
Cable length: 5m |
| Detection limit: 0.01 mg / L (20 ºC) |
Measurement limit: 40 mg / L |
| Response time: 2min (90%, 20 ºC) |
Polarization time: 60min |
| Minimum flow rate: 2.5cm / s |
Drift: <2% / month |
| Measurement error: <± 0.01 mg / L |
| Output Current: 50-80nA/0.1 mg / L Note: Maximum current 3.5uA |
| Polarization voltage: 0.7V |
Zero oxygen: <0.01 mg / L |
| Calibration interval:> 60 days |
Measured water temperature :0-60 ºC |
DOG-209F Industrial Dissolved Oxygen Sensor
Features
DOG-209F Dissolved Oxygen electrode has high stability and reliability, which can be used in the harsh environment;
it demands for less maintenance; it is suitable for continuous measurement of the dissolved oxygen in the fields of urban
sewage treatment, industrial waste water treatment, aquaculture, environmental monitoring etc.
Technical Indexes
| Measuring range: 0-20mg/L |
| Measuring principle: Current sensor (Polarographic electrode) |
| Permeable membrane thickness: 50 um |
| Electrode shell material: U PVC or 31 6L stainless steel |
| Temperature compensation resistor: Ptl00, Ptl000, 22K, 2.252K etc. |
| Sensor life: >2 years |
| Cable length: 5m |
| Detection lower limit: 0.01 mg/L (20ºC) |
| Measurement upper limit: 40mg/L |
| Response time: 3min (90%, 20ºC) |
| Polarization time: 60min |
| Minimum flow rate: 2.5cm/s |
| Drift: <2%/month |
| Measurement error: <± 0.1mg/I |
| Output current: 50~80nA/0.1mg/L Note: Maximum Current 3.5uA |
| Polarization voltage: 0.7V |
| Zero oxygen: <0.1 mg / L (5min) |
| Calibration intervals: >60 days |
| Measured water temperature: 0-60ºC |