Customization: | Available |
---|---|
Current Output: | 4-20mA |
Measuring Range: | 0-10% HCl;0-10%Naoh;0-10%H2so4;0-20%Nacl;0-10%Hno3 |
Still deciding? Get samples of US$ 400/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
1. Measuring range: concentration: (0.00~10.00)%HCL; (0.00~10.00)%NaOH, (0.00~30.00)%H2SO4; (0.00~10.00)%NaCL (bigger range customized), (0.00~15.00)%HNO3; temperature: 0~99.9ºC (Option) |
2. Intrinsic error of the electronic unit: concentration: ±1.0%, temperature: ±0.3ºC |
3. Automatic temperature compensation range:: 0~99.9ºC |
4. Solution tested: 0~99.9ºC, 0.6MPa |
5. Intrinsic error of the instrument: concentration: ±2.0%(the error may be less than 0.05% after usual point |
calibration) |
Temperature:±0.5ºC(0.0ºC≤T≤80.0ºC) |
6. Repeatability error of the electronic units: ±1.0% |
7. Stability of the electronic unit: ±1.0%/24h |
8. Isolated current output: 4~20 mA(load <750Ω) |
9. Output current error: ≤±1%FS |
10. Compensation range of the temperature coefficient: (0~10.0)%/ºC(adjustable) |
11. Alarm relay: AC220V, 3A |
12. RS485 Communication Interface |
13. Power supply: AC220V±22V, 50Hz±1Hz,24V DC(option) |
14. Protection grade: lP54, Aluminium shell, applicable for outdoor use. |
15. Clock accuracy: ±1 minute/month |
16. Data storage capacity: l month (1 point/5 minutes) |
17. Saving time of data under a continuous power-failure condition: 10 years |
18. Overall dimension: 146 (length) x 146 (width) x 150 (depth) mm; dimension of the hole: 138 x 138mm |
19. Weight: 1.5kg |
20. Working conditions: ambient temperature: 0-60ºC; relative humidity <85% |
Intelligent: SJG-3083 English On-line Acid Alkali Concentration Meter adopts high-precision AD conversion and |
single chip microcomputer processing technologies. It can be used for the measurement of the HCl, NaOH |
concentration and temperature, automatic temperature compensation, automatic range switching and |
self-checking. |
Reliability: All the components are arranged on one circuit board. No complicated functional switch, adjusting |
knob or potentiometer is arranged on this instrument. |
Strong interference immunity: Its current output adopts optoelectronic isolating technology. This meter has |
strong interference immunity and the capacity of long-distance transmission as well as excellent electromagnetic |
compatibility. |
Waterproof and dust-proof design: Its protection grade is IP54. It is applicable for outdoor use. |
Automatic range switching: The measuring ranges can automatically switch within the measuring range of the |
electrode. |
Communication interface: RS485, It can be easily connected to a computer to perform monitoring and communication. |
Phase sensitive detection: Eliminating the influence of conductors on the measurement of hydrochloric acid alkali |
concentration. |
Automatic measuring frequency switching: This can help prevent electrodes from polarization, so as to enhance |
their accuracy. |
FAQ
Q: Are you trading company or manufacturer ?
A: We are factory.
Q: How long is your delivery time?
A: Generally it is 5-10 days if the goods are in stock. or it is 15-20 days if the goods are not in stock,
it is according to quantity.
Q: Do you provide samples ? is it free or extra ?
A: We can offer the sample, it is extra
Q: What is your terms of payment ?
A: 100% T/T in advance ,balance before shippment.