Digital Pen Type Portable pH Meter (pH-2)

Product Details
Customization: Available
Display: Digital
Portable: Portable
Still deciding? Get samples of US$ 8/Piece
Request Sample
Gold Member Since 2014

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Number of Employees
111
Year of Establishment
2007-05-23
Find Similar Products

Basic Info.

Model NO.
PH-2
Measure Range
0.00pH-14.0 pH
Resolution
0.01pH
Accuracy
+/-0.01pH
Power Supply
2*1.5V
Operating Temperature
0oc--80oc
Transport Package
Carton Box
Specification
SGS, CE, ISO9001
Trademark
BOQU
Origin
Shanghai, China
HS Code
90278099
Production Capacity
10000piece/Month

Product Description

Technical Indexes
*measure range:               0.00pH-14.0 pH
*resolution:                          0.01ph
*accuracy:                        +/-0.01ph
*Power supply:           2*1.5V ( keep using more than 500 hours)
*operating temperature:            0ºC--80ºC
*calibration: two points automatic calibration 
*dimension: 153mm*29mm*16mm
*weight: 52g

Operation
1.remove  the  protective cap.
2.first rinse the electrode with distilled water, and suck it with filter paper.
3.turn the meter on by pressing the "ON" key.
4.immerse the ph meter electrode in the solution to be tested(can not be over the immersion line )
5.stir gently and wait around 30 seconds till  the reading stabilized.
6.after finished ,clear the electrode with pure water ,turn the meter off  by  pressing the "ON/OFF" key.
7.Always replace the protective cap after use.
 
 Calibration.
1. Turn on  pH meter
2.Immerse the electrode in pH 6.86 standard buffer solution.
   (under the temperature of 25°C)
3. Press the "CAL" button.
  Wait until the display digital is 6.86pH Then complete pH 6.86 calibration
  Clean the electrode with distilled water and dry it with filter paper. 
4.Immerse the electrode in pH 4.00 standard buffer solution, 
  Press "CAL" button.
  Wait until the digital is 4.00 pH..Complete the pH 4.00 point calibration.
5.Clean electrode and dry it , put pH meter into pH 9.18 standard buffer solution.
 Check if reading is correct between 9.17-9.19pH.
 
 
 
Attention
The instrument need recalibrate in the following condition.
1.Long time no use or use very long time after calibration
2.The electrode is used Very frequently.
3. The testing accuracy requirement is very high.
4.Press the "CAL" button long time in the air carelessly.
 
pH  electrode  maintenance

If the electrode placed long time, and electrode exposed to air, then showing values is slow and  not stable, because of electrode dry. You should immerse the electrode into water for few hours.
 
LowBatt 
When the display value is fuzzy or  unshown , The battery should be replaced  promptly.Pay attention to the polarity of battery.
 
Warranty
The instrument is warranted to the purchaser for a  year from the date of purchase.
If during these periods, repair or replacement of the instrument is required,
please return the instrument to either your dealer or to our office.The repair will be effected free of charge. ( damage is not due to negligence or erroneous operation by the user)
Note :Plese provide purchase invioce or proof before returning your instruments back .
 
 
 
 
 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier