DOG-208FA electrode, which is specially designed to be resistant to 130 degrees steam sterilization, the pressure auto-balance high temperature dissolved oxygen electrode, for liquids or gases dissolved oxygen measurement, the electrode is most suitable for small microbial culture reactor dissolved oxygen levels on line. Can also be used for environmental monitoring, waste water treatment and aquaculture on-line measurement dissolved oxygen levels.
1.Electrode body material: stainless steel |
2.Permeable membrane: fluorine plastic, |
silicone, stainless steel wire mesh composite membrane. |
3.Cathode: platinum wire |
4.Anode: silver |
5.Electrodes built-in temperature sensor: PT1000 |
6.The response current in the air: About 60nA |
7.The response current in a nitrogen atmosphere: less than |
one percent response current of response in air. |
8.Electrode response time: about 60 seconds (up 95% response) |
9.Electrode Response Stability: a constant oxygen partial pressure |
at a constant temperature environment, response current drift less |
than 3% per week |
Liquid mixing flow to the electrode response: 3% or less (in water at |
room temperature) |
Electrode Response Temperature Coefficient: 3% (greenhouse) |
Insert the electrode diameter: 12 mm, 19 mm, 25 mm optional |
Electrode insertion length: 80,150,200,250,300 mm |
1.Who are we?
BOQU instrument focus on R&D and production of Water Quality analyzer&sensor since 2007.
Our mission is To be brightest eye for water quality monitoring on earth.
★ Employees: 200+ people
★ Annual growth rate: 35%
★ R&D experiences: 20+ years
★ Technical patents: 50+
★ Annual production quantities: 150,000pcs
★ Cooperation companies: BOSCH, Boehringer Ingelheim, CM, Roche
★ Main Industries: Power plant, Water treatment plant, Drinking water, Pharmaceutical, Fish farming, Swimming pool.