pH-2 Digital pH Tester with High Quality Portable Type

Measure Range: 0.00pH-14.0 pH
Resolution: 0.01pH
Accuracy: +/-0.01pH
Power Supply: 2*1.5V
Operating Temperature: 0oc--80oc
Weight: 52g
Samples:
US$ 10/Piece 1 Piece(Min.Order)
| Request Sample
Customization:
Gold Member Since 2014

Suppliers with verified business licenses

Manufacturer/Factory

Basic Info.

Model NO.
PH-2
Dimensions
153*29*16
Transport Package
Carton Box
Specification
SGS, CE, ISO9001
Trademark
BOQU
Origin
Shanghai, China
HS Code
90278099
Production Capacity
10000piece/Month

Product Description

 Pocket-Size pH Meter pH Controller&pH meter (pH-2)

Technical Indexes
*measure range:               0.00pH-14.0 pH
*resolution:                          0.01ph
*accuracy:                        +/-0.01ph
*Power supply:           2*1.5V ( keep using more than 500 hours)
*operating temperature:            0ºC--80ºC
*calibration: two points automatic calibration 
*dimension: 153mm*29mm*16mm
*weight: 52g

Operation
1.remove  the  protective cap.
2.first rinse the electrode with distilled water, and suck it with filter paper.
3.turn the meter on by pressing the "ON" key.
4.immerse the ph meter electrode in the solution to be tested(can not be over the immersion line )
5.stir gently and wait around 30 seconds till  the reading stabilized.
6.after finished ,clear the electrode with pure water ,turn the meter off  by  pressing the "ON/OFF" key.
7.Always replace the protective cap after use. 

Calibration.
1. Turn on  pH meter
2.Immerse the electrode in pH 6.86 standard buffer solution.
   (under the temperature of 25°C)
3. Press the "CAL" button.
  Wait until the display digital is 6.86pH Then complete pH 6.86 calibration
  Clean the electrode with distilled water and dry it with filter paper. 
4.Immerse the electrode in pH 4.00 standard buffer solution, 
  Press "CAL" button.
  Wait until the digital is 4.00 pH..Complete the pH 4.00 point calibration.
5.Clean electrode and dry it , put pH meter into pH 9.18 standard buffer solution.
 Check if reading is correct between 9.17-9.19pH.
 
 Attention
The instrument need recalibrate in the following condition.
1.Long time no use or use very long time after calibration
2.The electrode is used Very frequently.
3. The testing accuracy requirement is very high.
4.Press the "CAL" button long time in the air carelessly.


FAQ
Q: Are you trading company or manufacturer ?
A: We are factory.

Q: How long is your delivery time?
A: Generally it is 5-10 days if the goods are in stock. or it is 15-20 days
if the goods are not in stock, it is according to quantity.

Q: Do you provide samples ? is it free or extra ?
A: We can offer the sample, it is extra

Q: What is your terms of payment ?
A: 100% T/T in advance ,balance before shippment.


If you have another question, pls feel free to contact us as below:
Contact: Louise Zhou
Tele: +86 021 33897320-8042
 
Mobile: +86 15180184494
 
 
 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now